Product Safety Recalls

Product Recall Tracker

AB Sciex, LLC Recall 92810

Description: Citrine Triple Quad Mass Spectrometer, Part Number 5063684

AB Sciex, LLC Recall 92810 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-0040-2024
Event ID92810
Event DescriptionCitrine Triple Quad Mass Spectrometer, Part Number 5063684
Product TypeDevices
DistributionUS Nationwide.
Quantity14
Recall ReasonIn rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.
Device Classification20231011
Device Code InfoUDI-DI: 00842365100533; Serial Numbers: DE220001803PL, DE230011804, DE230031806, DE230041807, DE230051808, DE230071901, DE230081903, DE230091903, DE230162003, DE230272105, DE230292105, DE240002106, DE240022111, DE240042203
Center Classification Date20231005
Recall Initiation Date20230830
Recalling FirmAB Sciex, LLC
Initial Notification Letter
Similar To

© 2022 Recall.Report / Contact / Privacy Policy