AB Sciex, LLC Recall 92810
Description: Citrine Triple Quad Mass Spectrometer, Part Number 5063684
AB Sciex, LLC Recall 92810 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-0040-2024 |
Event ID | 92810 |
Event Description | Citrine Triple Quad Mass Spectrometer, Part Number 5063684 |
Product Type | Devices |
Distribution | US Nationwide. |
Quantity | 14 |
Recall Reason | In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results. |
Device Classification | 20231011 |
Device Code Info | UDI-DI: 00842365100533; Serial Numbers: DE220001803PL, DE230011804, DE230031806, DE230041807, DE230051808, DE230071901, DE230081903, DE230091903, DE230162003, DE230272105, DE230292105, DE240002106, DE240022111, DE240042203 |
Center Classification Date | 20231005 |
Recall Initiation Date | 20230830 |
Recalling Firm | AB Sciex, LLC |
Initial Notification | Letter |
Similar To |