Philips Ultrasound Inc Recall 88625
Description: Philips Ultrasound, Inc. Sparq, Diagnostic Ultrasound System, REF 795116, 100-240V -, 50/60 Hz, 500 VA, CE2797, Rx Only,
Philips Ultrasound Inc Recall 88625 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-0071-2022 |
Event ID | 88625 |
Event Description | Philips Ultrasound, Inc. Sparq, Diagnostic Ultrasound System, REF 795116, 100-240V -, 50/60 Hz, 500 VA, CE2797, Rx Only, |
Product Type | Devices |
Distribution | Worldwide distribution - US Nationwide distribution in the states of CA, MA, MD, MI, MS, NC, NY, OH, VA, WA and the countries of Australia, Portugal, Netherlands, Canada, Netherlands, Germany. |
Quantity | 1 unit |
Recall Reason | Battery system data issue with the ultrasound system can intermittently cause a system shutdown, regardless of actual battery state or application of alternating current power. |
Device Classification | 20211013 |
Device Code Info | Affected Serial Number/UDI numbers: US92010594 (01)00884838098015(21)US92010594 |
Center Classification Date | 20211007 |
Recall Initiation Date | 20210819 |
Recalling Firm | Philips Ultrasound Inc |
Initial Notification | Letter |
Similar To |