Philips Ultrasound Inc Recall 88625

Description: Philips Ultrasound, Inc. Sparq, Diagnostic Ultrasound System, REF 795116, 100-240V -, 50/60 Hz, 500 VA, CE2797, Rx Only,

Philips Ultrasound Inc Recall 88625 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-0071-2022
Event ID88625
Event DescriptionPhilips Ultrasound, Inc. Sparq, Diagnostic Ultrasound System, REF 795116, 100-240V -, 50/60 Hz, 500 VA, CE2797, Rx Only,
Product TypeDevices
DistributionWorldwide distribution - US Nationwide distribution in the states of CA, MA, MD, MI, MS, NC, NY, OH, VA, WA and the countries of Australia, Portugal, Netherlands, Canada, Netherlands, Germany.
Quantity1 unit
Recall ReasonBattery system data issue with the ultrasound system can intermittently cause a system shutdown, regardless of actual battery state or application of alternating current power.
Device Classification20211013
Device Code InfoAffected Serial Number/UDI numbers: US92010594 (01)00884838098015(21)US92010594
Center Classification Date20211007
Recall Initiation Date20210819
Recalling FirmPhilips Ultrasound Inc