Product Safety Recalls

Product Recall Tracker

LEICA BIOSYSTEMS NUSSLOCH GMBH Recall 95323

Description: Leica Biosystems Cryostat Model Number CM1900, Material Numbers: a) 14045227382, b) 14901900100, c) 14901900700, d) 1491900UVU2; for freezing and sectioning tissue samples

LEICA BIOSYSTEMS NUSSLOCH GMBH Recall 95323 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-0088-2025
Event ID95323
Event DescriptionLeica Biosystems Cryostat Model Number CM1900, Material Numbers: a) 14045227382, b) 14901900100, c) 14901900700, d) 1491900UVU2; for freezing and sectioning tissue samples
Product TypeDevices
DistributionUS Nationwide distribution.
Quantity25 units
Recall ReasonAs part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potentially ignite and therefore can cause serious injuries. The flammability is indicated on the labelling of these sprays. The cryostat labelling, however, did not include any specific warning about usage of flammable freezing sprays prior to 2019.
Device Classification20241023
Device Code Infoa) Material Number 14045227382, UDI/DI 04049188075840, Serial Numbers: 1198121999, 383051996; b) Material Number 14901900100, UDI/DI 04049188003324, Serial Numbers: 703111997, 1018021999, 1538042001, 2058, 629071997, 2994, 1170111999, 3686102007, 3698; c) Material Number 14901900700, UDI/DI 04049188003300, Serial Numbers: 1267032000, 2602062004, 1577052001, 1679092001, 2542042004, 1771, 1885032002, 2682092004, 2376112003, 2774122004, 1114, 1749112001, 1997072002; d) Material Number 1491900UVU2, UDI/DI 04049188002570, Serial Numbers: 3243052006
Center Classification Date20241015
Recall Initiation Date20240904
Recalling FirmLEICA BIOSYSTEMS NUSSLOCH GMBH