Siemens Healthcare Diagnostics, Inc. Recall 90872
Description: Dimension Vista High Density Lipoprotein Cholesterol (HDLC) Flex reagent cartridge, IVD Siemens Material Number (SMN): 10464340 Catalog Number: K3048A
Siemens Healthcare Diagnostics, Inc. Recall 90872 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-0089-2023 |
Event ID | 90872 |
Event Description | Dimension Vista High Density Lipoprotein Cholesterol (HDLC) Flex reagent cartridge, IVD Siemens Material Number (SMN): 10464340 Catalog Number: K3048A |
Product Type | Devices |
Distribution | Nationwide Foreign: Canada |
Quantity | 1592 units |
Recall Reason | Dimension Vista¿ High Density Lipoprotein Cholesterol (HDLC) Potential for Erroneous Result. If an HDLC QC or patient result is obtained without an Abnormal Reaction [E145]: flag, a falsely depressed or elevated result ranging from -90% to 133% may be observed. |
Device Classification | 20221026 |
Device Code Info | UDI-DI : 0084276802043821250BA2022090710464340840 Lot Number: 21250BA Exp. Date: 2022-09-07 |
Center Classification Date | 20221014 |
Recall Initiation Date | 20220824 |
Recalling Firm | Siemens Healthcare Diagnostics, Inc. |
Initial Notification | Letter |
Similar To |