Product Safety Recalls

Product Recall Tracker

Siemens Healthcare Diagnostics, Inc. Recall 90872

Description: Dimension Vista High Density Lipoprotein Cholesterol (HDLC) Flex reagent cartridge, IVD Siemens Material Number (SMN): 10464340 Catalog Number: K3048A

Siemens Healthcare Diagnostics, Inc. Recall 90872 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-0089-2023
Event ID90872
Event DescriptionDimension Vista High Density Lipoprotein Cholesterol (HDLC) Flex reagent cartridge, IVD Siemens Material Number (SMN): 10464340 Catalog Number: K3048A
Product TypeDevices
DistributionNationwide Foreign: Canada
Quantity1592 units
Recall ReasonDimension Vista¿ High Density Lipoprotein Cholesterol (HDLC) Potential for Erroneous Result. If an HDLC QC or patient result is obtained without an Abnormal Reaction [E145]: flag, a falsely depressed or elevated result ranging from -90% to 133% may be observed.
Device Classification20221026
Device Code InfoUDI-DI : 0084276802043821250BA2022090710464340840 Lot Number: 21250BA Exp. Date: 2022-09-07
Center Classification Date20221014
Recall Initiation Date20220824
Recalling FirmSiemens Healthcare Diagnostics, Inc.
Initial Notification Letter
Similar To

© 2022 Recall.Report / Contact / Privacy Policy

This site's data is aggregated programmatically and provided "as is" without any representations or warranties, express or implied.
Recall.report is not affiliated with the Consumer Product Safety Commision, FDA, or US government in any way.