SIEMENS HEALTHCARE DIAGNOSTICS INC Recall 90867

Description: N Antiserum to Human IgG, 2mL Variant for BN II System and BN ProSpec System-IVD Siemens Material Number (SMN): 10446297

SIEMENS HEALTHCARE DIAGNOSTICS INC Recall 90867 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-0093-2023
Event ID90867
Event DescriptionN Antiserum to Human IgG, 2mL Variant for BN II System and BN ProSpec System-IVD Siemens Material Number (SMN): 10446297
Product TypeDevices
DistributionNationwide
Quantity373 PCK
Recall ReasonN Antiserum to Human IgG are not meeting the current High-Dose Hook Effect expectation of up to 648 mg/L for IgGU (urine) and up to 2290 mg/L for IgGC (CSF).
Device Classification20221026
Device Code InfoUDI-DI: 0405686900179VF Lot Number: 153088 Future lots may be impacted by the issue but will then be provided with a corresponding instruction note within the reagent package
Center Classification Date20221014
Recall Initiation Date20220830
Recalling FirmSIEMENS HEALTHCARE DIAGNOSTICS INC
Initial Notification Letter