TECHNO-PATH MANUFACTURING LTD. Recall 93017

Description: Multichem P, Part Code 08P90-10

TECHNO-PATH MANUFACTURING LTD. Recall 93017 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-0106-2024
Event ID93017
Event DescriptionMultichem P, Part Code 08P90-10
Product TypeDevices
DistributionUS Nationwide distribution in the state of Illinois.
Quantity2964 units
Recall ReasonTechnopath Manufacturing Ltd. has identified an issue with the glass vial commodity used to fill the Multichem S Plus Level 1, Multichem S Plus Level 2, Multichem S Plus Level 3, Multichem IA Plus and Multichem P Controls. Some customers have experienced broken vials when receiving or thawing Technopath Multichem IA Plus and Multichem S Plus Controls. Breakages have been observed upon delivery, while thawing or while opening the vial. The breakages can result in the bottom of the vial breaking off, breakage in the neck thread area or cracking on the sides of the vial. This may expose the user to biohazardous material if cut and may cause a delay in patient results.
Device Classification20231018
Device Code InfoUDI/DI 05391523440769, Lot Numbers: 04530621P, 04561021P, 04610523P
Center Classification Date20231012
Recall Initiation Date20230825
Recalling FirmTECHNO-PATH MANUFACTURING LTD.
Initial Notification E-Mail
Similar To