Fort Defiance Industries, LLC Recall 88628
Description: P2131 Automated Field Steam Sterilizer NSN: 6530-01-641-4641
Fort Defiance Industries, LLC Recall 88628 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-0121-2022 |
Event ID | 88628 |
Event Description | P2131 Automated Field Steam Sterilizer NSN: 6530-01-641-4641 |
Product Type | Devices |
Distribution | Distributed nationwide to PA, MD, TX, NJ, OH, TN, UT, VA and internationally to Canada. |
Quantity | 380 devices |
Recall Reason | The rear heater plate and cover assembly that contains the electrical wiring for the heaters does not have sufficient bonding to ground. |
Device Classification | 20211027 |
Device Code Info | Model number: P2131; NSN: 6530-01-641-4641; Serial Numbers AFS-0001 through AFS-0386. Note: AFS-0297 was not manufactured. |
Center Classification Date | 20211015 |
Recall Initiation Date | 20210819 |
Recalling Firm | Fort Defiance Industries, LLC |