Product Safety Recalls

Product Recall Tracker

Fort Defiance Industries, LLC Recall 88628

Description: P2131 Automated Field Steam Sterilizer NSN: 6530-01-641-4641

Fort Defiance Industries, LLC Recall 88628 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-0121-2022
Event ID88628
Event DescriptionP2131 Automated Field Steam Sterilizer NSN: 6530-01-641-4641
Product TypeDevices
DistributionDistributed nationwide to PA, MD, TX, NJ, OH, TN, UT, VA and internationally to Canada.
Quantity380 devices
Recall ReasonThe rear heater plate and cover assembly that contains the electrical wiring for the heaters does not have sufficient bonding to ground.
Device Classification20211027
Device Code InfoModel number: P2131; NSN: 6530-01-641-4641; Serial Numbers AFS-0001 through AFS-0386. Note: AFS-0297 was not manufactured.
Center Classification Date20211015
Recall Initiation Date20210819
Recalling FirmFort Defiance Industries, LLC
Initial Notification E-Mail
Similar To

© 2022 Recall.Report / Contact / Privacy Policy

This site's data is aggregated programmatically and provided "as is" without any representations or warranties, express or implied.
Recall.report is not affiliated with the Consumer Product Safety Commision, FDA, or US government in any way.