Implant Resource Inc Recall 88585
Description: MPR Slotted Mallet, labeling etched onto devices as follows: Medical Products Resource 800.524.5194 www.m-p-r.com MPR-1120-1000 (01)00840001405868
Implant Resource Inc Recall 88585 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-0122-2022 |
Event ID | 88585 |
Event Description | MPR Slotted Mallet, labeling etched onto devices as follows: Medical Products Resource 800.524.5194 www.m-p-r.com MPR-1120-1000 (01)00840001405868 |
Product Type | Devices |
Distribution | US distribution to Florida and New Hampshire |
Quantity | 6 units |
Recall Reason | Potential for mallet head to become separated from mallet handle. There is also significant deformations on striking surface that were not expected |
Device Classification | 20211027 |
Device Code Info | Model Number: MPR-1120-1000, Lot Number: 200911-1-1 |
Center Classification Date | 20211015 |
Recall Initiation Date | 20210721 |
Recalling Firm | Implant Resource Inc |
Initial Notification | |
Similar To |