Product Safety Recalls

Product Recall Tracker

Insulet Corporation Recall 90987

Description: 18239: ASM Omnipod Dash PDM, insulin delivery system. PT-000010: Assembly, Dash Final PDM U100, mg/dL PT-000011: Assembly, Dash Final PDM U100, mmol/L PT-000030: Assembly, DASH PDM, Canada Remote Controller (Personal Diabetes Manager or PDM) associated with Insulin Pump (Pod) *The part numbers above may be incorporated into higher level assemblies such as DASH Starter Kits.

Insulet Corporation Recall 90987 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-0154-2023
Event ID90987
Event Description18239: ASM Omnipod Dash PDM, insulin delivery system. PT-000010: Assembly, Dash Final PDM U100, mg/dL PT-000011: Assembly, Dash Final PDM U100, mmol/L PT-000030: Assembly, DASH PDM, Canada Remote Controller (Personal Diabetes Manager or PDM) associated with Insulin Pump (Pod) *The part numbers above may be incorporated into higher level assemblies such as DASH Starter Kits.
Product TypeDevices
DistributionWorldwide Distribution: US nationwide, Canada, Europe, Australia, and the Middle East.
Quantity248,288 US; 167,802 OUS
Recall ReasonThe firm has become aware of PDM battery issues, including battery swelling, fluid leaking from the battery, and extreme overheating, which may pose a fire hazard. The firm has received a complaint alleging fire as a result of thermal energy from a DASH PDM.
Device Classification20221123
Device Code InfoPart Number/UDI-DI 18239 : 10385082000009 PT 000030: 10385082000122 PT 000010: 10385082000139 PT 000011: 10385082000146 All Serial Numbers and Lots
Center Classification Date20221115
Recall Initiation Date20221017
Recalling FirmInsulet Corporation
Initial Notification Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Similar To

© 2022 Recall.Report / Contact / Privacy Policy

This site's data is aggregated programmatically and provided "as is" without any representations or warranties, express or implied.
Recall.report is not affiliated with the Consumer Product Safety Commision, FDA, or US government in any way.