Omega Medical Imaging, Inc. Recall 93018
Description: Omega Medical Imaging Elevating Monitor Suspension, Part Number 1000-0085
Omega Medical Imaging, Inc. Recall 93018 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-0156-2024 |
Event ID | 93018 |
Event Description | Omega Medical Imaging Elevating Monitor Suspension, Part Number 1000-0085 |
Product Type | Devices |
Distribution | Worldwide distribution - US Nationwide and the country of Pakistan. |
Quantity | 52 units |
Recall Reason | An actuator separated from the pivot mechanism on a lower monitor boom system column and became detached and was restrained by the installed safety cable at one location. |
Device Classification | 20231101 |
Device Code Info | all units manufactured between 10-22-2012 and 07-09-2019 |
Center Classification Date | 20231024 |
Recall Initiation Date | 20230228 |
Recalling Firm | Omega Medical Imaging, Inc. |
Initial Notification | |
Similar To |