Omega Medical Imaging, Inc. Recall 93018

Description: Omega Medical Imaging Elevating Monitor Suspension, Part Number 1000-0085

Omega Medical Imaging, Inc. Recall 93018 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-0156-2024
Event ID93018
Event DescriptionOmega Medical Imaging Elevating Monitor Suspension, Part Number 1000-0085
Product TypeDevices
DistributionWorldwide distribution - US Nationwide and the country of Pakistan.
Quantity52 units
Recall ReasonAn actuator separated from the pivot mechanism on a lower monitor boom system column and became detached and was restrained by the installed safety cable at one location.
Device Classification20231101
Device Code Infoall units manufactured between 10-22-2012 and 07-09-2019
Center Classification Date20231024
Recall Initiation Date20230228
Recalling FirmOmega Medical Imaging, Inc.