AB SCIEX Recall 88748
Description: Cliquid MD version 3.4 software used in conjunction with Analyst MD software on SCIEX Citrine & 4500MD series liquid chromatography-tandem mass spectrometry (LC-MS/MS) systems. Model No. 5088288
AB SCIEX Recall 88748 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-0175-2022 |
Event ID | 88748 |
Event Description | Cliquid MD version 3.4 software used in conjunction with Analyst MD software on SCIEX Citrine & 4500MD series liquid chromatography-tandem mass spectrometry (LC-MS/MS) systems. Model No. 5088288 |
Product Type | Devices |
Distribution | Worldwide distribution - US Nationwide distribution in the states of District of Columbia, Florida, Illinois, Louisiana, Maryland, Michigan, Missouri, New Jersey, New York, North Carolina, Ohio, Pennsylvania, Tennessee, Utah, Wisconsin and the countries of Austria, Finland, Germany, Hong Kong, Italy, Japan, Norway, Spain, Switzerland, United Kingdom, Italy, France, and Japan. |
Quantity | 30 |
Recall Reason | The values of the Internal Standard (IS) concentrations are incorrectly derived when the user builds a customized test, then set the IS column to hidden and left the column empty. This may lead to the user making incorrect conclusions using incorrect results. |
Device Classification | 20211103 |
Device Code Info | Software version 3.4 |
Center Classification Date | 20211026 |
Recall Initiation Date | 20211004 |
Recalling Firm | AB SCIEX |
Initial Notification | Letter |
Similar To |