3M Company – Health Care Business Recall 88763
Description: 3M Red Dot Electrodes, Radiolucent Monitoring Electrode with Foam Tape, Catalog No. 2244, (3M ID (SKU): 70-2011-8506-6, SAP ID: 7100212989), packaged 10 electrodes per card, 5 cards per pouch, 20 pouches per shipper box, 1,000 electrodes total per shipper box. The 3M Red Dot Radiolucent Monitoring Electrode with Foam Tape 2244 is intended to be used by healthcare professionals for ECG monitoring. This electrode is disposable, intended for single use, and has been tested for up to 3 days wear.
3M Company – Health Care Business Recall 88763 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-0188-2022 |
Event ID | 88763 |
Event Description | 3M Red Dot Electrodes, Radiolucent Monitoring Electrode with Foam Tape, Catalog No. 2244, (3M ID (SKU): 70-2011-8506-6, SAP ID: 7100212989), packaged 10 electrodes per card, 5 cards per pouch, 20 pouches per shipper box, 1,000 electrodes total per shipper box. The 3M Red Dot Radiolucent Monitoring Electrode with Foam Tape 2244 is intended to be used by healthcare professionals for ECG monitoring. This electrode is disposable, intended for single use, and has been tested for up to 3 days wear. |
Product Type | Devices |
Distribution | US Nationwide Distribution to states of: AL, AZ, CA, IA, IL, LA, MD, MO, NC, NJ, and OH; and OUS Foreign distribution to countries of: Australia, Bermuda, Canada, China, France, Germany, Japan, the Netherlands, Singapore, Sweden, and Turkey. |
Quantity | 2,081,000 electrodes |
Recall Reason | Potential for breakage of the ECG electrode carbon stud from the eyelet. |
Device Classification | 20211110 |
Device Code Info | Lot #202302LK, Exp. 2023-02-19, pouch UDI (01)1070738789449(17)230219(11)202302LK, shipper box UDI (01)50707387789447(17)230219(11)202302LK Lot #202302LL, Exp. 2023-02-22, pouch UDI (01)1070738789449(17)230222(11)202302LL, shipper box UDI (01)50707387789447(17)230222(11)202302LL; Lot #202302LM, Exp. 2023-02-22, pouch UDI (01)1070738789449(17)230222(11)202302LM, shipper box UDI (01)50707387789447(17)230222(11)202302LM; Lot #202302LN, Exp. 2023-02-23. pouch UDI (01)1070738789449(17)230223(11)202302LN, shipper box UDI (01)50707387789447(17)230223(11)202302LN. |
Center Classification Date | 20211101 |
Recall Initiation Date | 20211006 |
Recalling Firm | 3M Company - Health Care Business |
Initial Notification | Letter |
Similar To |