Jjgc Industria E Comercio De Materials Dentarios Sa Recall 88917
Description: GM Helix Acqua Implant 3.5x10, Article No. 140.944
Jjgc Industria E Comercio De Materials Dentarios Sa Recall 88917 Information
Status | Terminated |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-0203-2022 |
Event ID | 88917 |
Event Description | GM Helix Acqua Implant 3.5x10, Article No. 140.944 |
Product Type | Devices |
Distribution | US Nationwide distribution. |
Quantity | 567 (US) |
Recall Reason | 3.5x11.5mm implants were mislabeled as 3.5x10mm implants. In borderline cases where the bone quality is low and the bone ridge has a maximum indication for the implant height (10mm), if the incorrect implant is installed, it could result in damage to the nerve or adjacent bone structures/cavities. |
Device Classification | 20211110 |
Device Code Info | UDI 7899878024897 Lots GNW15 and GRE82 |
Center Classification Date | 20211104 |
Recall Initiation Date | 20211011 |
Recalling Firm | Jjgc Industria E Comercio De Materials Dentarios Sa |
Initial Notification | Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit |
Similar To |