X-NAV Technologies, LLC Recall 91067

Description: X-Guide X-Mark Probe Tool -used for registering edentulous patients to the software or performing XMark registration. Catalog Number:P010673

X-NAV Technologies, LLC Recall 91067 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-0222-2023
Event ID91067
Event DescriptionX-Guide X-Mark Probe Tool -used for registering edentulous patients to the software or performing XMark registration. Catalog Number:P010673
Product TypeDevices
DistributionUS Nationwide distribution and Foreign country of: Canada
Quantity102 units
Recall ReasonProduct missing one of two welds connecting the tip to the shaft, if the remaining weld breaks, it is possible that the tip could fall out during X-Mark registration and fall into the patient s mouth. In this case, the patient may aspirate or swallow the probe tip.
Device Classification20221116
Device Code InfoUDI-DI: (01)00817421020995 (10) 2007010010 Lot number: 2007010010
Center Classification Date20221110
Recall Initiation Date20220928
Recalling FirmX-NAV Technologies, LLC
Initial Notification Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit