Becton Dickinson & Co. Recall 91010

Description: BD Phoenix PMIC/ID-107: The Phoenix PMIC/ID panel is used for the rapid identification and susceptibility of most aerobic and facultative anaerobic gram-positive bacteria of human origin with selected antimicrobial agents when used with the BD Phoenix Automated Microbiology System. Catalog (Ref) Number: 448607

Becton Dickinson & Co. Recall 91010 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-0232-2023
Event ID91010
Event DescriptionBD Phoenix PMIC/ID-107: The Phoenix PMIC/ID panel is used for the rapid identification and susceptibility of most aerobic and facultative anaerobic gram-positive bacteria of human origin with selected antimicrobial agents when used with the BD Phoenix Automated Microbiology System. Catalog (Ref) Number: 448607
Product TypeDevices
DistributionUS Nationwide distribution.
Quantity13,900 EA
Recall ReasonMisformulated minocycline-can cause erroneously increased Minimal Inhibitory Concentration (MIC) values when performing antibiotic susceptibility testing (AST) against minocycline and result in the reporting of false resistance for clinical samples if the initial Quality Control (QC) testing of the panel batch passes, delaying treatment of the patient with the most appropriate antibiotic
Device Classification20221123
Device Code InfoUDI: (10) 1355046 (17) 230131 (30) 0025 (01) 30382904486076 Lot Number: 1355046 Exp. Date: 31JAN2023
Center Classification Date20221114
Recall Initiation Date20221013
Recalling FirmBecton Dickinson & Co.
Initial Notification Letter