Product Safety Recalls

Product Recall Tracker

Medicalplastic S.R.L. Recall 88785

Description: Ilizarov Wire Tensioner, Model 71070341, UDI (10)08033201842072(10)21HP02550(11)210831(21E434

Medicalplastic S.R.L. Recall 88785 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-0245-2022
Event ID88785
Event DescriptionIlizarov Wire Tensioner, Model 71070341, UDI (10)08033201842072(10)21HP02550(11)210831(21E434
Product TypeDevices
DistributionMemphis, TN
Quantity581 units
Recall ReasonMedicalplastic s.r.l. is conducting a voluntary field action to remove a group of Ilizarov Wire Tensioners due to a manufacturing error. An internal washer of the Ilizarov Wire Tensioner has been made in a wrong plastic material that could wear and break during the sterilization process in autoclave. Consequently this could cause malfunctioning of the tensioner.
Device Classification20211124
Device Code InfoLot # 20LP02813, Serial #D147 to D149; Lot # 20LP02812, Serial # D115 to D119 ; Lot # 20LP02816, Serial # D099 to D114 ; Lot # 20HP01926, Serial # D059 to D065 ; Lot # 20LP02819, Serial # D175 to D198 ; Lot # 20LP02818, Serial # D129 to D146 ; Lot # 20LP02815, Serial # D120 to D128 ; Lot # 20MP03078, Serial # D374 ; Lot # 20HP01927, Serial # D278 to D282 ; Lot # 20MP03037, Serial # D294 to D296 ; Lot # 20LP02827, Serial # D199 to D258 ; Lot # 20 MP03036, Serial # D288 to D293 ; Lot # 20MP03077, Serial # D370 to D373 ; Lot # 20MP03035, Serial # D259 to D277 ; Lot # 20MP03041, Serial # D283 to D287 ; Lot # 20MP03049, Serial # D343 to D369 ; Lot # 20MP03048, Serial # D297 to D342 ; Lot # 20MP03123, Serial # D375 to D459 ; Lot # 20HP01928, Serial # D460 to D474 ; Lot # 20MP03126, Serial # D482 to D491 ; Lot # 20MP03125, Serial # D492 to D522 ; Lot # 20MP03124, Serial # D475 to D481 ; Lot # 21BP00265, Serial # D523 to D574 ; Lot # 21BP00266, Serial # D626 to D672 ; Lot # 21BP00271, Serial # D588 to D625 ; Lot # 21BP00272, Serial # D575 to D587 ;
Center Classification Date20211117
Recall Initiation Date20210617
Recalling FirmMedicalplastic S.R.L.
Initial Notification E-Mail
Similar To

© 2022 Recall.Report / Contact / Privacy Policy

This site's data is aggregated programmatically and provided "as is" without any representations or warranties, express or implied.
Recall.report is not affiliated with the Consumer Product Safety Commision, FDA, or US government in any way.