The Binding Site Group, Ltd. Recall 88921

Description: Freelite Human Lambda Free Kit (for use on the SPAPLUS) REF LK018.S Freelite Human Lambda Free Kit (for use on the SPAPLUS) (x10) REF LK018.10S

The Binding Site Group, Ltd. Recall 88921 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-0274-2022
Event ID88921
Event DescriptionFreelite Human Lambda Free Kit (for use on the SPAPLUS) REF LK018.S Freelite Human Lambda Free Kit (for use on the SPAPLUS) (x10) REF LK018.10S
Product TypeDevices
DistributionU.S.: CA, CO, FL, IL, IN, KY, LA, MI, NY, OK, SD, TX and VA. O.U.S.: Canada
Quantity2,254 kits
Recall ReasonDue to calibration curves producing a higher activity than routinely expected and that there may be a decrease in separation between calibrator point 5 and calibrator point 6, which can be exacerbated with calibrator 6 activity reaches the optical ceiling of the SPA+ analyzer.
Device Classification20211201
Device Code InfoModel Number: LK018.S UDI Code: 05051700006477 Lot Number: 472776 Model Number:LK018.10S UDI Code: 05051700012201 Lot Number: 472664
Center Classification Date20211119
Recall Initiation Date20211015
Recalling FirmThe Binding Site Group, Ltd.