BioFire Diagnostics, LLC Recall 88934

Description: BioFire Blood Culture Identification 2 (BCID2) Panel Part No: RFIT-ASY-0147 UDI: 00815381020338 Blood Culture Catalog No./Description: 442023/BD BACTEC Plus Aerobic medium in plastic vials 442021/BD BACTEC Lytic Anaerobic medium in plastic vials 442192//BD BACTEC Plus Aerobic medium in glass vials 442022/BD BACTEC Plus Anaerobic medium in plastic vials 442020/BD BACTEC Peds Plus mediums in plastic vials All lots with an expiry date prior to

BioFire Diagnostics, LLC Recall 88934 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-0312-2022
Event ID88934
Event DescriptionBioFire Blood Culture Identification 2 (BCID2) Panel Part No: RFIT-ASY-0147 UDI: 00815381020338 Blood Culture Catalog No./Description: 442023/BD BACTEC Plus Aerobic medium in plastic vials 442021/BD BACTEC Lytic Anaerobic medium in plastic vials 442192//BD BACTEC Plus Aerobic medium in glass vials 442022/BD BACTEC Plus Anaerobic medium in plastic vials 442020/BD BACTEC Peds Plus mediums in plastic vials All lots with an expiry date prior to
Product TypeDevices
DistributionWorldwide distribution - U.S. Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and WY. The Countries of Algeria, Argentina, Australia, Austria, Bahrain, Belarus, Belgium, Bulgaria, Chile, Colombia, Costa Rica, Croatia, Czech Republic, Dominican Rep/, Ecuador, Egypt, France, Georgia, Germany, Greece, Guam, Guatemala, Hong Kong, Hungary, India, Israel Italy, Kuwait, Latvia, Macedonia, Malaysia, Mexico, Netherlands, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Serbia, Shanghai, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, United Kingdom, Uruguay, Utd/Arab Emir/.
QuantityTotal 14,140 Kits [11,155 kits (U.S.) and 2,985 kits (O.U.S.)]
Recall ReasonDue to an increased risk of false positive Acinetobacter calcoaceticus- baumannii complex results. The cause may be as a due to non-viable contamination in the blood culture vials.
Device Classification20211208
Device Code InfoBlood Culture Identification 2 (BCID2) Panel Part Number: RFIT-ASY-0147 UDI Code: 00815381020338 Blood Culture Media Catalog No./Description: 442023/BD BACTEC Plus Aerobic medium in plastic vials 442021/BD BACTEC Lytic Anaerobic medium in plastic vials 442192//BD BACTEC Plus Aerobic medium in glass vials 442022/BD BACTEC Plus Anaerobic medium in plastic vials 442020/BD BACTEC Peds Plus mediums in plastic vials All BCID2 Panel lots BD Blood Culture Media vials with expiration date of April 30, 2022 or earlier.
Center Classification Date20211130
Recall Initiation Date20211011
Recalling FirmBioFire Diagnostics, LLC