Randox Laboratories Ltd. Recall 88999
Description: RANDOX CHOL-cholesterol (total) test system intended to measure cholesterol in plasma and serum REF: CH200
Randox Laboratories Ltd. Recall 88999 Information
Status | Terminated |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-0328-2022 |
Event ID | 88999 |
Event Description | RANDOX CHOL-cholesterol (total) test system intended to measure cholesterol in plasma and serum REF: CH200 |
Product Type | Devices |
Distribution | US Nationwide distribution in the states of GA, OR, PA ,WA, West Virginia. |
Quantity | 15 kits |
Recall Reason | When calibrating using the standard provided in the kit, the change in absorbance is lower than expected generating internal quality control results high outside range |
Device Classification | 20211208 |
Device Code Info | Lot Numbers/Exp Date: 559344 28th June 2022; 572730 28th July 2023. GTIN: 05055273201130 |
Center Classification Date | 20211201 |
Recall Initiation Date | 20211101 |
Recalling Firm | Randox Laboratories Ltd. |