Instrumentation Laboratory Recall 93313
Description: GEM Premier 3000 PAK - pH, pCO2, pO2, Hct, Na+, K+, Ca++, Part Numbers 00024307507, 00024315007, 00024330007, 00024345007, 00024360007
Instrumentation Laboratory Recall 93313 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-0358-2024 |
Event ID | 93313 |
Event Description | GEM Premier 3000 PAK - pH, pCO2, pO2, Hct, Na+, K+, Ca++, Part Numbers 00024307507, 00024315007, 00024330007, 00024345007, 00024360007 |
Product Type | Devices |
Distribution | US States: MA, MN, NY. India. |
Quantity | 2448 units |
Recall Reason | Internal testing identified that the GEM Premier PAKs (cartridges) may exhibit an over-recovery (i.e., positive bias) of pO2 results in the low range (at pO2 levels less than 60 mmHg) that are outside of labeled performance specifications for pO2. The positive bias may lead to false positive for medical decision levels at pO2 < 60 mmHg, which may result in failure to detect and/or adequately treat hypoxemia and lead to a delay in treatment. The quality control procedures for the Non-iQM GEM Premier 3000 may not detect the non-conformance. |
Device Classification | 20231129 |
Device Code Info | Part No. 00024307507: UDI-DI: 08426950087946, Lot Numbers: 323031H, 323531D, 324231M, 325031D, 324931F, 323831E, 323031I, 324231I, 324831E, 325631C; Part No. 00024315007: UDI-DI: 08426950082484, Lot Numbers: 323441B, 323641G, 323041H, 323541D, 324141J, 324841E, 325041D, 325641C; Part No. 00024330007: UDI-DI: 08426950082514, Lot Numbers: 323051H, 323551D, 323651G, 324151J; Part No. 00024345007: UDI-DI: 08426950082545, Lot Numbers: 323761H; Part No. 00024360007: UDI-DI: 08426950082569, Lot Numbers: 323771H, 323471B, 324871E; |
Center Classification Date | 20231120 |
Recall Initiation Date | 20230929 |
Recalling Firm | Instrumentation Laboratory |
Initial Notification | Letter |