Instrumentation Laboratory Recall 93313

Description: GEM Premier 3000 PAK - pH, pCO2, pO2, Hct, Na+, K+, Ca++, Part Numbers 00024307507, 00024315007, 00024330007, 00024345007, 00024360007

Instrumentation Laboratory Recall 93313 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-0358-2024
Event ID93313
Event DescriptionGEM Premier 3000 PAK - pH, pCO2, pO2, Hct, Na+, K+, Ca++, Part Numbers 00024307507, 00024315007, 00024330007, 00024345007, 00024360007
Product TypeDevices
DistributionUS States: MA, MN, NY. India.
Quantity2448 units
Recall ReasonInternal testing identified that the GEM Premier PAKs (cartridges) may exhibit an over-recovery (i.e., positive bias) of pO2 results in the low range (at pO2 levels less than 60 mmHg) that are outside of labeled performance specifications for pO2. The positive bias may lead to false positive for medical decision levels at pO2 < 60 mmHg, which may result in failure to detect and/or adequately treat hypoxemia and lead to a delay in treatment. The quality control procedures for the Non-iQM GEM Premier 3000 may not detect the non-conformance.
Device Classification20231129
Device Code InfoPart No. 00024307507: UDI-DI: 08426950087946, Lot Numbers: 323031H, 323531D, 324231M, 325031D, 324931F, 323831E, 323031I, 324231I, 324831E, 325631C; Part No. 00024315007: UDI-DI: 08426950082484, Lot Numbers: 323441B, 323641G, 323041H, 323541D, 324141J, 324841E, 325041D, 325641C; Part No. 00024330007: UDI-DI: 08426950082514, Lot Numbers: 323051H, 323551D, 323651G, 324151J; Part No. 00024345007: UDI-DI: 08426950082545, Lot Numbers: 323761H; Part No. 00024360007: UDI-DI: 08426950082569, Lot Numbers: 323771H, 323471B, 324871E;
Center Classification Date20231120
Recall Initiation Date20230929
Recalling FirmInstrumentation Laboratory