Smith & Nephew, Inc. Recall 89031
Description: smith&nephew EVOS MEDIAL DISTAL FEMUR PLATE, R, 5H, 90MM, REF 72573201
Smith & Nephew, Inc. Recall 89031 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-0364-2022 |
Event ID | 89031 |
Event Description | smith&nephew EVOS MEDIAL DISTAL FEMUR PLATE, R, 5H, 90MM, REF 72573201 |
Product Type | Devices |
Distribution | US Nationwide distribution in the states of CO, FL, GA, MA, MN, MO, NC, NJ, NV, NY, OH, OK, PA, TN, TX, WA. |
Quantity | 35 units |
Recall Reason | Testing vendor notified Smith & Nephew that previous MRI scanning conditions provided in Exponent s reports were erroneous. Erroneous data from the reports was incorporated in the released IFU 81110294 EVOS Large Fragment Plating System. |
Device Classification | 20211222 |
Device Code Info | Batch Number: 20HM10255, 20HM10256, 20HM10257, UDI |
Center Classification Date | 20211210 |
Recall Initiation Date | 20211101 |
Recalling Firm | Smith & Nephew, Inc. |
Initial Notification | Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit |
Similar To |