Carefusion 2200 Inc Recall 90967

Description: CareFusion V. Mueller TITANIUM BAYONET BIPOLAR FORCEPS, 0.5MM TIP, ANGLED-UP FINE OVERALL LENGTH 8-3/4" (225MM), REF F-5127

Carefusion 2200 Inc Recall 90967 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-0411-2023
Event ID90967
Event DescriptionCareFusion V. Mueller TITANIUM BAYONET BIPOLAR FORCEPS, 0.5MM TIP, ANGLED-UP FINE OVERALL LENGTH 8-3/4" (225MM), REF F-5127
Product TypeDevices
DistributionNationwide, Austria, Canada, Egypt, Germany, Hong Kong, Israel, Philippines, Saudi Arabia, South Africa, South Korea, Thailand
Quantity5 units
Recall ReasonBD has determined that the Instructions For Use (IFU) provided by BD is missing certain content that is contained in the manufacturer's IFU. This missing content is related to the interface with the device power supply, as well as cleaning and maintenance instructions. Refer to the attached V. Mueller" Bipolar Forcep IFU.
Device Classification20221214
Device Code InfoUDI/DI 10885403041235, ALL LOTS
Center Classification Date20221203
Recall Initiation Date20221102
Recalling FirmCarefusion 2200 Inc