Medtronic Navigation, Inc. Recall 89058
Description: Stealthstation System w/ Stealthstation Cranial Software 3.1.1,3.1.2,3.1. or SynergyCranial Software. Synergy Cranial, Model: 9733763, and StealthStation Cranial, Model: 9735585
Medtronic Navigation, Inc. Recall 89058 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-0383-2022 |
Event ID | 89058 |
Event Description | Stealthstation System w/ Stealthstation Cranial Software 3.1.1,3.1.2,3.1. or SynergyCranial Software. Synergy Cranial, Model: 9733763, and StealthStation Cranial, Model: 9735585 |
Product Type | Devices |
Distribution | US: MN, FL, MD, IL, AK, NE, NH, PA, ND, AR, CA, WI, NJ, GA, SD, AZ, KY, AL, MS, MO, TX, MI, MT, OH, NY, VA, MA, CT, TN, WV, WA, NC, CO, NV, ME, LA, DE, NM, IN, UT, DC, OR, SC, OK, IA, HI, ID, KS, WY, MN, PR OUS: AU, BN, GU, JP, KR, MY, NZ, SG, TH, VN, MM, CA, CZ, RU, AL, CY, DE, ES, FR, GB, HU, IL, IT, NL, PL, RO, CN, HK, TW, IN, LK, NP, AR, BB, BR, CL, CO, CR, DO, EC, MX, PE, PR, VE, NI, UY, AE, BH, DZ, EG, IQ, JO, KZ, KE, SA, LB, LY, MA, NA, PK, QA, ZA, TR, AT, BE, CH, DK, FI, GF, GR, IE, LV, LT, LU, PT, SE, SK, UK, BA, BG, CI, HR, RS, SI |
Quantity | 2587 |
Recall Reason | Cranial biopsy procedure software can enter a state where the biopsy depth gauge is no longer synchronized with the rest of the navigational information on the screen and may display an incorrect position of the biopsy needle, which may result in prolonged procedure, the need for additional surgical procedure, tissue injury. |
Device Classification | 20211229 |
Device Code Info | 9733763, Version: 2.2.8; 9735585, Versions: 3.1.1, 3.1.2, 3.1.3 |
Center Classification Date | 20211222 |
Recall Initiation Date | 20211111 |
Recalling Firm | Medtronic Navigation, Inc. |
Initial Notification | Letter |
Similar To |