Cook Inc. Recall 89104

Description: Skinny Needle with Chiba Tip. RPN 090010, 090010-ET; GPN G14254, G15290. Rigid needle available with EchoTip to provide enhanced visualization of needle tip when used with ultrasonic imaging equipment.

Cook Inc. Recall 89104 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-0414-2022
Event ID89104
Event DescriptionSkinny Needle with Chiba Tip. RPN 090010, 090010-ET; GPN G14254, G15290. Rigid needle available with EchoTip to provide enhanced visualization of needle tip when used with ultrasonic imaging equipment.
Product TypeDevices
DistributionWorldwide distribution - US Nationwide distribution in the states of AL, CA, IN, LA, MS, MT, NC, NY, OR, PA, WA and the countries of France, Italy, and Mexico.
Quantity112 US, 240 OUS
Recall ReasonSpecific lots may contain black debris on the exterior of the inner stylet and/or on the interior of the outer cannula. Local inflammatory reactions may occur if an affected product is used.
Device Classification20211229
Device Code InfoRPN 090010/GPN G14254: LOT NUMBER UDI 13977925 (01)00827002142541(17)240524(10)13977925 14030949 (01)00827002142541(17)240617(10)14030949 14054162 (01)00827002142541(17)240628(10)14054162 14116047 (01)00827002142541(17)240728(10)14116047 RPN 090010-ET/GPN G15290: LOT NUMBER UDI 14038615 (01)00827002152901(17)240618(10)14038615
Center Classification Date20211222
Recall Initiation Date20211013
Recalling FirmCook Inc.
Initial Notification Letter
Similar To