Product Safety Recalls

Product Recall Tracker

Medtronic Navigation, Inc. Recall 89074

Description: Stealth Autoguide Tracker, Model: 28248, which is an instrument in the Stealth Autoguide Basic Instrument Kit, Mode: 9736188, and the Bundle Autoguide Reg Kit.

Medtronic Navigation, Inc. Recall 89074 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-0417-2022
Event ID89074
Event DescriptionStealth Autoguide Tracker, Model: 28248, which is an instrument in the Stealth Autoguide Basic Instrument Kit, Mode: 9736188, and the Bundle Autoguide Reg Kit.
Product TypeDevices
DistributionUS: FL, NE, CA, MN, TX, OH, IL, AZ, KS, KY, OR, OK, WA, NV, MO, IN, MI, UT, MS, VA, SC. OUS: Canada, Sweden, Italy, France, Germany, Spain, Czech Republic, United Kingdom, India, Nepal, Japan, Turkey, China
Quantity98
Recall ReasonA tracker that holds and guides neurosurgical surgical instruments may be subject to separation/weld failure, where the tracker connects to the instrument tube. If a weld separation occurs and a navigation inaccuracy is experienced, it may result in prolonged procedure and tissue injury. Per 11/24/21 firm email: The Bundle Autoguide Reg Kit does not have a model number. The Bundle Autoguide Reg Kit is internal terminology and not visible to the customer and are used as sales and shipping configuration only. When this sales bundle is chosen by an internal Medtronic Sales Representative the devices included in the bundle are individually packaged and labeled and sent to the customer in a single shipment. The devices included in the bundle are ordered, packed, and shipped as individual medical devices with unique model numbers, which is the reason that the Autoguide Tracker (28248) and the Stealth Autoguide Basic Instrument Kit (9736188) are the only part numbers subject to this FCA.
Device Classification20211229
Device Code InfoAll lot numbers
Center Classification Date20211222
Recall Initiation Date20211109
Recalling FirmMedtronic Navigation, Inc.
Initial Notification Letter
Similar To

© 2022 Recall.Report / Contact / Privacy Policy

This site's data is aggregated programmatically and provided "as is" without any representations or warranties, express or implied.
Recall.report is not affiliated with the Consumer Product Safety Commision, FDA, or US government in any way.