Aizu Olympus Co., Ltd. Recall 93335

Description: PCF-H190L, EVIS EXERA III COLONOVIDEOSCOPE

Aizu Olympus Co., Ltd. Recall 93335 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-0419-2024
Event ID93335
Event DescriptionPCF-H190L, EVIS EXERA III COLONOVIDEOSCOPE
Product TypeDevices
DistributionDomestic distribution to FL, KY, MA, NJ, NY, and PA
Quantity24 incorrectly repaired units
Recall ReasonSpecific devices may have an improperly repaired light guide connector due to missing adhesive to secure the suction channel mouthpiece to the scope connector. Issue may result in loss of image, decreased suction, or impact the ability to effectively reprocess the device. This in turn may lead to unanticipated delay to treatment, or patient exposure to contaminated material and subsequent infection.
Device Classification20231206
Device Code InfoUDI-DI 04953170305191 Serial Numbers 2302719, 2516751
Center Classification Date20231129
Recall Initiation Date20231016
Recalling FirmAizu Olympus Co., Ltd.