Karl Storz Endoscopy Recall 88998
Description: Karl Storz Udel Sterilization Tray Instructions for Use, Tray Part Number/Description: 39301ACTS/CAMERA TRAY (STERRAD, INLINE CAMERA HEADS); 39301AS/RIGID TELESCOPE TRAY (STERRAD, 2 RIGID SCOPES 4MM X 18CM); 39301BCTS/CAMERA TRAY (INLINE/C-MOUNT CAMERA HEADS); 39301BS/RIGID TELESCOPE TRAY (STERRAD, 2 RIGID SCOPES 5MM X 35CM); 39301C1S/RIGID TELESCOPE TRAY (STERRAD, 1 RIGID SCOPE & LT CABLE); 39301CS/RIGID TELESCOPE TRAY (STERRAD, 2 RIGID SCOPE 10MM X 31CM); 39301DS/RIGID TELESCOPE TRAY (STERRAD, 2 RIGID SCOPES); 39301HCTS/CAMERA TRAY (STERRAD, STANDARD CAMERA HEADS); 39301PHTS/CAMERA TRAY (STERRAD, PENDULUM/INLINE CAM HEADS); 39305C1S/BASKET-STYLE RIGID TELESCOPE TRAY (UDEL, 1 RIGID SCOPE 4MM X 30CM); 39305C2S/BASKET-STYLE RIGID TELESCOPE TRAY (STERRAD, 2 RIGID SCOPES); 39305L1S/BASKET-STYLE RIGID TELESCOPE TRAY (STERRAD, 1 RIGID SCOPE); 39305L2S/BASKET-STYLE RIGID TELESCOPE TRAY (STERRAD, 2 RIGID SCOPES); 39311AS/RIGID TELESCOPE TRAY (STERRAD, 4 RIGID SCOPES & LT CABLE); 39311BS/RIGID TELESCOPE TRAY (STERRAD, TWO LAP SCOPES/CABLE); 39314FS/RIGID TELESCOPE TRAY (STERRAD, FOUR LAP SCOPES/CABLE); 39401AS/FLEXIBLE ENDOSCOPE TRAY (UP TO 39CM); 39402AS/FLEXIBLE ENDOSCOPE TRAY (FIBERSCOPES); 39403AS/FLEXIBLE ENDOSCOPE TRAY (CCD VIDESCOPES); 39406AS/FLEXIBLE ENDOSCOPE TRAY (CMOS VIDEOSCOPES)
Karl Storz Endoscopy Recall 88998 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-0442-2022 |
Event ID | 88998 |
Event Description | Karl Storz Udel Sterilization Tray Instructions for Use, Tray Part Number/Description: 39301ACTS/CAMERA TRAY (STERRAD, INLINE CAMERA HEADS); 39301AS/RIGID TELESCOPE TRAY (STERRAD, 2 RIGID SCOPES 4MM X 18CM); 39301BCTS/CAMERA TRAY (INLINE/C-MOUNT CAMERA HEADS); 39301BS/RIGID TELESCOPE TRAY (STERRAD, 2 RIGID SCOPES 5MM X 35CM); 39301C1S/RIGID TELESCOPE TRAY (STERRAD, 1 RIGID SCOPE & LT CABLE); 39301CS/RIGID TELESCOPE TRAY (STERRAD, 2 RIGID SCOPE 10MM X 31CM); 39301DS/RIGID TELESCOPE TRAY (STERRAD, 2 RIGID SCOPES); 39301HCTS/CAMERA TRAY (STERRAD, STANDARD CAMERA HEADS); 39301PHTS/CAMERA TRAY (STERRAD, PENDULUM/INLINE CAM HEADS); 39305C1S/BASKET-STYLE RIGID TELESCOPE TRAY (UDEL, 1 RIGID SCOPE 4MM X 30CM); 39305C2S/BASKET-STYLE RIGID TELESCOPE TRAY (STERRAD, 2 RIGID SCOPES); 39305L1S/BASKET-STYLE RIGID TELESCOPE TRAY (STERRAD, 1 RIGID SCOPE); 39305L2S/BASKET-STYLE RIGID TELESCOPE TRAY (STERRAD, 2 RIGID SCOPES); 39311AS/RIGID TELESCOPE TRAY (STERRAD, 4 RIGID SCOPES & LT CABLE); 39311BS/RIGID TELESCOPE TRAY (STERRAD, TWO LAP SCOPES/CABLE); 39314FS/RIGID TELESCOPE TRAY (STERRAD, FOUR LAP SCOPES/CABLE); 39401AS/FLEXIBLE ENDOSCOPE TRAY (UP TO 39CM); 39402AS/FLEXIBLE ENDOSCOPE TRAY (FIBERSCOPES); 39403AS/FLEXIBLE ENDOSCOPE TRAY (CCD VIDESCOPES); 39406AS/FLEXIBLE ENDOSCOPE TRAY (CMOS VIDEOSCOPES) |
Product Type | Devices |
Distribution | US: AK, AL, AP, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY OUS: Canada, Venezuela, Barbados, GUAM, Puerto Rico, Mexico, Northern Mariana Islands, Trinidad & Tobago, Germany, Cayman Islands, Philippines, Japan, Dominican Republic, Panama, Bahamas, Ecuador, Suriname, American Virgin Islands, British Virgin Islands, Belize, Curacao, The Bahamas, Brazil, South Korea, Bermuda |
Quantity | 71892 |
Recall Reason | To ensure sterilization effectiveness, sterilization tray instructions for use were updated and now state 1) udel trays may be used with flexible endoscopes with lumen diameters greater-than-or-equal to 1.2mm and maximum length of 845mm and 2) ethylene oxide is excluded as a sterilization modality. |
Device Classification | 20220105 |
Device Code Info | All lots with Instructions for Use Prior to PI-000078-10.0 (06/21) |
Center Classification Date | 20211229 |
Recall Initiation Date | 20211015 |
Recalling Firm | Karl Storz Endoscopy |
Initial Notification | Letter |
Similar To |