Product Safety Recalls

Product Recall Tracker

Elekta, Inc. Recall 93264

Description: ELEKTA LIMITED Elekta Synergy Platform, REF MRT 8501, MRT 8511; Medical Charged-Particle Radiation Therapy System Accelerator, Linear

Elekta, Inc. Recall 93264 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-0444-2024
Event ID93264
Event DescriptionELEKTA LIMITED Elekta Synergy Platform, REF MRT 8501, MRT 8511; Medical Charged-Particle Radiation Therapy System Accelerator, Linear
Product TypeDevices
Distributionworldwide
Quantity264 units
Recall ReasonElekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT / XVI detector arms. When the detector arm is fully deployed there is no risk of uncontrolled extension.
Device Classification20231213
Device Code InfoUDI/DI 05060191071536, Serial Numbers: 152110, 153550, 153363, 152650, 151851, 153300, 152721, 152816, 152539, 153879, 154099, 154121, 154230, 154581, 151909, 152186, 152314, 152542, 152728, 152643, 151984, 153334, 152057, 152086, 152107, 152129, 152907, 153327, 153820, 153844, 153084, 151712, 151713, 153522, 151873, 152501, 153675, 151834, 151923, 151924, 151925, 152076, 154100, 151725, 151804, 151825, 151826, 152337, 152338, 152339, 152662, 152936, 153629, 153632, 153676, 153678, 152397, 153636; UDI/DI 05060191071567, Serial Numbers: 152718, 153868, 153870, 153877, 153948, 154180, 152778, 151562, 151563, 151991, 151992, 151993, 152213, 152450, 152456, 152531, 152642, 152663, 152782, 152791, 152829, 152862, 152962, 153125, 153270, 153343, 153759, 153789, 153886, 154145, 154147, 154247, 154330, 154499, 153925, 154201, 152171, 152280, 151750, 152221, 152255, 152509, 152558, 152612, 152831, 153066, 153135, 153229, 153246, 153704, 153772, 154191, 154428, 154510, 152146, 152355, 151994, 152001, 153098, 152436, 152935, 153414, 153811, 154270, 154471, 154472, 154136, 151858, 152209, 152384, 152497, 152498, 153111, 153156, 153362, 153446, 153718, 154058, 152486, 153234, 153747, 151772, 151840, 151949, 152170, 152312, 152368, 152480, 152672, 152687, 152739, 152753, 152899, 152904, 152913, 152919, 152952, 152977, 153060, 153168, 153284, 153297, 153325, 153349, 153551, 153553, 153605, 153660, 153770, 153813, 153888, 153920, 153955, 154006, 154228, 154238, 154240, 154446, 109008, 151731, 152702, 152869, 154029, 154102, 151453, 152840, 152841, 153388, 153389, 153440, 153683, 152873, 152958, 153423, 151872, 153266, 153505, 153714, 151880, 152341, 152342, 153326, 153205, 153977, 153178, 152319, 154010, 152306, 153737, 153760, 152191, 152638, 153059, 154235, 154309, 151964, 153119, 154041, 152367, 153972, 153981, 153983, 153992, 152184, 152708, 152231, 153809, 151863, 152270, 152794, 152982, 153516, 153885, 154253, 154295, 154318, 154357, 151665, 152058, 152059, 152404, 152723, 152598, 154463, 152120, 152143, 153089, 153092, 153579, 153850, 154122, 152479, 152858, 152878, 151849, 152155, 152205, 152483, 152377, 152378, 152380, 152381, 153137, 154037, 154563, 153688,
Center Classification Date20231201
Recall Initiation Date20231025
Recalling FirmElekta, Inc.
Initial Notification E-Mail
Similar To

© 2022 Recall.Report / Contact / Privacy Policy

This site's data is aggregated programmatically and provided "as is" without any representations or warranties, express or implied.
Recall.report is not affiliated with the Consumer Product Safety Commision, FDA, or US government in any way.