Invacare Corporation Recall 89062
Description: Invacare AVIVA FX with LiNX Gyro Power Wheelchair (with LiNX Gyro Module Model Numbers DLX-GRY100-A and DLX-GRY101-A) Models: IFX-20R, IFX-20SP, IFX-20MP, IFX-20C
Invacare Corporation Recall 89062 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-0446-2022 |
Event ID | 89062 |
Event Description | Invacare AVIVA FX with LiNX Gyro Power Wheelchair (with LiNX Gyro Module Model Numbers DLX-GRY100-A and DLX-GRY101-A) Models: IFX-20R, IFX-20SP, IFX-20MP, IFX-20C |
Product Type | Devices |
Distribution | Worldwide distribution - US Nationwide and the countries of Canada, Australia, Thailand, New Zealand. |
Quantity | 303 units US; 20 units OUS |
Recall Reason | Power Wheelchairs with a LiNX Gyro component running firmware version 6.1.2 can experience a more aggressive deceleration rate than the programmed rate due to the system following the incorrect deceleration profile, resulting in potential injury |
Device Classification | 20220112 |
Device Code Info | LiNX Gyro firmware version 6.1.2 UDI-DI Numbers: IFX-20C 00841447111627 IFX-20MP 00841447111658 IFX-20R 00841447111634 IFX-20SP 00841447111641 IFX20WIDE 00841447111658 |
Center Classification Date | 20220103 |
Recall Initiation Date | 20211112 |
Recalling Firm | Invacare Corporation |
Initial Notification | Letter |
Similar To |