Product Safety Recalls

Product Recall Tracker

Flower Orthopedics Corporation Recall 89092

Description: Flex-Thread Inserter Shaft intended for Fibula intramedullary-intended to be used with the Inserter Draw Rod and Blue Ratchet Handle in order to facilitate the insertion of the Flex-Thread Nail System- Catalog Number: 8435-1

Flower Orthopedics Corporation Recall 89092 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-0448-2022
Event ID89092
Event DescriptionFlex-Thread Inserter Shaft intended for Fibula intramedullary-intended to be used with the Inserter Draw Rod and Blue Ratchet Handle in order to facilitate the insertion of the Flex-Thread Nail System- Catalog Number: 8435-1
Product TypeDevices
DistributionUS Nationwide distribution in the states of AZ, CA, CO, GA, TN, FL, IL, MA, MD, NC, TX.
Quantity48 units
Recall ReasonMating incompatibility between the Inserter Shaft and the Inserter Draw Rod causing the inserter draw rod to bind and not rotate freely as intended surgeon may be unable to complete surgery unless a second set is available
Device Classification20220112
Device Code InfoLot Number: 11090 UDI: 00840118117548
Center Classification Date20220103
Recall Initiation Date20211102
Recalling FirmFlower Orthopedics Corporation
Initial Notification Letter
Similar To

© 2022 Recall.Report / Contact / Privacy Policy