PTW NORTH AMERICA CORPORATION Recall 89314

Description: Software BeamAdjust version 2.2, Catalog No: S070009 Software VeriSoft version 8.0, Catalog No: S080032

PTW NORTH AMERICA CORPORATION Recall 89314 Information

StatusCompleted
Mandated?Voluntary: Firm initiated
Recall NumberZ-0479-2022
Event ID89314
Event DescriptionSoftware BeamAdjust version 2.2, Catalog No: S070009 Software VeriSoft version 8.0, Catalog No: S080032
Product TypeDevices
DistributionUS Nationwide distribution in the state of CO.
Quantity1 unit
Recall ReasonWhen a measurement with a PTW detector array is performed with the software BeamAdjust 2.2 or VeriSoft 8.0, a measurement error can occur under specific conditions. This error can lead to a too-low or too-high measured absolute dose which affects patient plan verification results
Device Classification20220119
Device Code InfoS/N :AU2001094-1 UDI: EPTWS0700090 VeriSoft ;EPTWS0700090 BeamAdjust
Center Classification Date20220111
Recall Initiation Date20210827
Recalling FirmPTW NORTH AMERICA CORPORATION
Initial Notification Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit