Covidien Llc Recall 89182
Description: LigaSure Blunt Tip Laparoscopic Sealer/Divider, Nano-Coated, Product Number LF1837
Covidien Llc Recall 89182 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-0495-2022 |
Event ID | 89182 |
Event Description | LigaSure Blunt Tip Laparoscopic Sealer/Divider, Nano-Coated, Product Number LF1837 |
Product Type | Devices |
Distribution | Worldwide Distribution in the countries of Australia, Austria, Belgium, Czech Republic, France, Germany, Ireland, Kazakhstan, Netherlands, Spain, United Kingdom. No US distribution. |
Quantity | 504 devices |
Recall Reason | Customers reported that the device jaws were difficult to open or would not open following application on tissue. Use of a device with this potential assembly issue could result in the potential for bleeding, unintended tissue injury or loss, and delay in treatment while an alternate device is obtained. |
Device Classification | 20220119 |
Device Code Info | UDI: 20884521705880 10884521705883; Lot Number: 11320187X |
Center Classification Date | 20220113 |
Recall Initiation Date | 20211206 |
Recalling Firm | Covidien Llc |
Initial Notification | Letter |
Similar To |