Siemens Medical Solutions USA, Inc Recall 89260

Description: Artis pheno with software VE20C- diagnostic imaging angiography system Model: 10849000

Siemens Medical Solutions USA, Inc Recall 89260 Information

StatusTerminated
Mandated?Voluntary: Firm initiated
Recall NumberZ-0515-2022
Event ID89260
Event DescriptionArtis pheno with software VE20C- diagnostic imaging angiography system Model: 10849000
Product TypeDevices
DistributionUS Nationwide distribution.
Quantity12 units
Recall ReasonFour potential software issues with Artis pheno and Artis icono systems with software VE20C: 1. Updated calibration data not saved with measurement after scene+/-; 2. No x-ray possible, system shutdown/restart might be required during intervention; 3. Corrupted Image during Roadmap; 4. Unintended shutdown of Imaging System with UPS (Uninterruptable Power Supply) option
Device Classification20220202
Device Code InfoSerial Numbers: 164719 164712 164722 164723 164728 164711 164718 164703 164721 164726 164724 164725 UDI: 04056869046877
Center Classification Date20220121
Recall Initiation Date20211012
Recalling FirmSiemens Medical Solutions USA, Inc