Immuno-Mycologics, Inc Recall 89370

Description: CRYPTOCOCCAL ANTIGEN LATERAL FLOW ASSAY, IMMY, Lateral Flow Assay for the Detection of Cryptococcal Antigen in serum and CSF, REF CR2003, IVD, UDI: (01)00816387020919

Immuno-Mycologics, Inc Recall 89370 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-0545-2022
Event ID89370
Event DescriptionCRYPTOCOCCAL ANTIGEN LATERAL FLOW ASSAY, IMMY, Lateral Flow Assay for the Detection of Cryptococcal Antigen in serum and CSF, REF CR2003, IVD, UDI: (01)00816387020919
Product TypeDevices
DistributionUS Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IL, IN, KY, LA, MA, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NY, OH, OR, PA, RI, SC, TN, TX, UT, VA, WA.
Quantity411 kits
Recall ReasonThe firm found immunoassay products used to detect cryptococcal antigen in serum and cerebrospinal fluid to have reduced specificity (90% now versus 99% before). As a result, a small number of samples with positive test results may be false positives, which may cause some patients to initiate unnecessary anti-fungal therapy. Samples with negative test results are NOT affected.
Device Classification20220209
Device Code InfoF1021035 and F1021036
Center Classification Date20220131
Recall Initiation Date20220103
Recalling FirmImmuno-Mycologics, Inc