Abiomed, Inc. Recall 89328

Description: OXY-1 System, Model Number 001-0500-001. Used during cardiopulmonary bypass. OXY-1 System consists of Console, Pump Drive, E Drive, and Pole Mount. The disposable Pump Lung Unit PLU part number 001-0400-100 is sold separately and is not part of this recall.

Abiomed, Inc. Recall 89328 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-0550-2022
Event ID89328
Event DescriptionOXY-1 System, Model Number 001-0500-001. Used during cardiopulmonary bypass. OXY-1 System consists of Console, Pump Drive, E Drive, and Pole Mount. The disposable Pump Lung Unit PLU part number 001-0400-100 is sold separately and is not part of this recall.
Product TypeDevices
DistributionUS distribution to FL, GA, IN, MD, NC, PA, TX.
Quantity42 Units
Recall ReasonThe firm has received complaints of the OXY-1 System Console experiencing a power interruption while in use, which disrupts therapy delivered to a patient on support. Disruption of therapy could lead to prolonged hypoxia.
Device Classification20220209
Device Code InfoUDI 00860001797909 Software Version Number V-1.3.15.0, V-1.4.6.0 and V.1.4.8.0 Serial Numbers 20200231-04 20200249-01 20200249-05 20200309-02 20200339-01 20200339-03 20200352-01 20200442-01 20200442-02 20200442-05 20210768-01 20210795-02 20210870-01 20210880-01 20210890-01 20210894-01 20210895-01 20210896-01 20210897-01 20210795-04
Center Classification Date20220203
Recall Initiation Date20211214
Recalling FirmAbiomed, Inc.
Initial Notification Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Similar To