Luminex Corporation Recall 89244
Description: ARIES HSV 1&2 Assay, Catalog/REF# 50-10017
Luminex Corporation Recall 89244 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-0553-2022 |
Event ID | 89244 |
Event Description | ARIES HSV 1&2 Assay, Catalog/REF# 50-10017 |
Product Type | Devices |
Distribution | Worldwide distribution - US Nationwide distribution in the states of New Jersey, California, Connecticut, Minnesota, Missouri, New Jersey, New Mexico, Pennsylvania, Tennessee, Washington, and Wisconsin) and the country of Malaysia. |
Quantity | 329 |
Recall Reason | Impacted lot may leak inside the ARIES instrument. Testing confirmed a leak rate of 0.9%. Leaking cassettes could potentially lead to incorrect results or a delay in reporting results due to an Invalid. |
Device Classification | 20220209 |
Device Code Info | Lot Numbers AB3600A and AB3560A |
Center Classification Date | 20220202 |
Recall Initiation Date | 20211117 |
Recalling Firm | Luminex Corporation |
Initial Notification | |
Similar To |