Baxter Healthcare Corporation Recall 89327
Description: Baxter Equipo de transferencia para DP MiniCap de vida util prolongada con pinza de torsion, 5C4482S
Baxter Healthcare Corporation Recall 89327 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-0574-2022 |
Event ID | 89327 |
Event Description | Baxter Equipo de transferencia para DP MiniCap de vida util prolongada con pinza de torsion, 5C4482S |
Product Type | Devices |
Distribution | Worldwide |
Quantity | 116,628 units |
Recall Reason | Cleaning products, containing, but not limited to, hydrogen peroxide, bleach, alcohol, or antiseptic agents and solvent, such as those containing acetone, toluene, xylene, or cyclohexanone may cause damage, such as leaking or cracking, to the MiniCap Extended Life PD transfer sets. |
Device Classification | 20220216 |
Device Code Info | All lot numbers within expiry |
Center Classification Date | 20220208 |
Recall Initiation Date | 20220107 |
Recalling Firm | Baxter Healthcare Corporation |
Initial Notification | Letter |
Similar To |