Baxter Healthcare Corporation Recall 89327
Description: Baxter CAPD System Transfer Set (CAPD MiniCap Transfer Set II (34cm)), Code No. T5C4484
Baxter Healthcare Corporation Recall 89327 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-0580-2022 |
Event ID | 89327 |
Event Description | Baxter CAPD System Transfer Set (CAPD MiniCap Transfer Set II (34cm)), Code No. T5C4484 |
Product Type | Devices |
Distribution | Worldwide |
Quantity | 2,240 units |
Recall Reason | Cleaning products, containing, but not limited to, hydrogen peroxide, bleach, alcohol, or antiseptic agents and solvent, such as those containing acetone, toluene, xylene, or cyclohexanone may cause damage, such as leaking or cracking, to the MiniCap Extended Life PD transfer sets. |
Device Classification | 20220216 |
Device Code Info | All lot numbers within expiry |
Center Classification Date | 20220208 |
Recall Initiation Date | 20220107 |
Recalling Firm | Baxter Healthcare Corporation |
Initial Notification | Letter |
Similar To |