Insulet Corporation Recall 93511

Description: Omnipod 5 App (on compatible Android smartphones), Software Versions 1.1 - 1.2.3 (current). US: PT-000599- Omnipod 5 App Software Package (provided through Google Play), EU: PT-001323- Omnipod 5 App Software Package (provided on locked-down Controllers (PT-001194 and PT-001193). Used to control an insulin pump.

Insulet Corporation Recall 93511 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-0591-2024
Event ID93511
Event DescriptionOmnipod 5 App (on compatible Android smartphones), Software Versions 1.1 - 1.2.3 (current). US: PT-000599- Omnipod 5 App Software Package (provided through Google Play), EU: PT-001323- Omnipod 5 App Software Package (provided on locked-down Controllers (PT-001194 and PT-001193). Used to control an insulin pump.
Product TypeDevices
DistributionWorldwide distribution - US Nationwide and the countries of Germany and UK.
Quantity27000 users
Recall ReasonThe bolus calculator is not recording the decimal point if it is the first value entered when changing a bolus dose. This may lead to an over-delivery of insulin to the user if the user does not recognize the error on the bolus calculator screen or the confirmation screen prior to starting the bolus.
Device Classification20240110
Device Code InfoUDI/Model/NDC Number: " Omnipod 5 App Software UDI: 1038508112030 " EU locked-down Controller UDI: 10385083000190 and 10385083000206
Center Classification Date20240104
Recall Initiation Date20231130
Recalling FirmInsulet Corporation
Initial Notification Letter
Similar To

© 2022 Recall.Report / Contact / Privacy Policy

This site's data is aggregated programmatically and provided "as is" without any representations or warranties, express or implied.
Recall.report is not affiliated with the Consumer Product Safety Commision, FDA, or US government in any way.