GE Healthcare (China) Co., Ltd. Recall 95704

Description: GE Healthcare Revolution Eagle, Model Number 5849504-3, X-ray/computed tomography system

GE Healthcare (China) Co., Ltd. Recall 95704 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-0607-2025
Event ID95704
Event DescriptionGE Healthcare Revolution Eagle, Model Number 5849504-3, X-ray/computed tomography system
Product TypeDevices
DistributionWorldwide distribution.
Quantity4 units
Recall ReasonGE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical, cine, and cardiac (if applicable) scan modes, after the first image of an exam, each subsequent image can be rotated by an increasing amount. The amount of progressive image rotation is dependent on the gantry speed and the duration of the scan. The final image in the exam can be rotated up to a maximum of ~56 degrees. The entire anatomy rotates as a whole by equal degrees, without distortion, and the anatomical positioning and z-axis of the images remains correct. Small degrees of image rotation might not be noticed. If image rotation is noticed and necessitates a rescan of the patient, the same image rotation will reoccur.
Device Classification20241211
Device Code InfoUDI/DI 00195278385215, Serial/Sales Order Numbers: CBDPG2400006HM, CBDPG2400005HM, CBDPG2400004HM, CBDPG2400007HM
Center Classification Date20241204
Recall Initiation Date20241105
Recalling FirmGE Healthcare (China) Co., Ltd.