Enztec Limited Recall 84217
Description: REF 800-05-035, ENZTEC FEMORAL IMPACTOR - DJO, (01) 09421028110624 (10) 999999, To be used only with DJO EMPOWR 3D KNEE
Enztec Limited Recall 84217 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-0612-2020 |
Event ID | 84217 |
Event Description | REF 800-05-035, ENZTEC FEMORAL IMPACTOR - DJO, (01) 09421028110624 (10) 999999, To be used only with DJO EMPOWR 3D KNEE |
Product Type | Devices |
Distribution | CA |
Quantity | 312 Femoral Impactors |
Recall Reason | The femoral impactors have a higher than normal risk of the arm breaking. During impaction the tip of the arm may fracture and separate from the instrument. This may cause a minor delay in surgery. If not retrieved, it may cause an inflammatory response that may require revision surgery. |
Device Classification | 20210203 |
Device Code Info | Lot number Distribution Date: 161471 08-Jun-16, 161472 08-Jun-16, 161473 08-Jun-16, 161474 08-Jun-16, 161475 08-Jun-16, 161476 08-Jun-16, 161477 08-Jun-16, 161756 29-Jun-16, 161757 29-Jun-16, 161758 29-Jun-16, 161759 29-Jun-16, 161760 29-Jun-16, 161761 29-Jun-16, 161762 29-Jun-16, 161822 22-Jul-16, 161823 22-Jul-16, 161824 22-Jul-16, 161825 22-Jul-16, 161826 22-Jul-16, 161827 22-Jul-16, 161828 22-Jul-16, 170760 24-Mar-17, 170761 24-Mar-17, 170762 24-Mar-17, 170763 24-Mar-17, 170764 24-Mar-17, 170765 24-Mar-17, 170766 24-Mar-17, 170767 24-Mar-17, 170768 24-Mar-17, 171989 23-Jun-17 , |
Center Classification Date | 20191206 |
Recall Initiation Date | 20190816 |
Recalling Firm | Enztec Limited |