Landauer Recall 93524
Description: Landauer microSTARii reader, Medical Dosimetry System, Model Numbers: 18000-000/18001-000/18007-000
Landauer Recall 93524 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-0616-2024 |
Event ID | 93524 |
Event Description | Landauer microSTARii reader, Medical Dosimetry System, Model Numbers: 18000-000/18001-000/18007-000 |
Product Type | Devices |
Distribution | US and Australia, Brazil, Canada, China, France, Germany, India, Japan, Mexico, Rep of Korea, Taiwan, Thailand, Viet Nam |
Quantity | 584 units |
Recall Reason | MicroSTARii readers may be contributing to the nanodot dosimeter measurement inaccuracy in two possible ways. The first is related to the interaction of the reader s LED beam profile with the nanoDot dosimeter. The second is debris formation due to drawer actuation, resulting in a change of reader response that may not be detectable unless known dose QCs are being run regularly. |
Device Classification | 20240103 |
Device Code Info | All serial numbers. |
Center Classification Date | 20231222 |
Recall Initiation Date | 20231115 |
Recalling Firm | Landauer |