Landauer Recall 93524

Description: Landauer microSTARii reader, Medical Dosimetry System, Model Numbers: 18000-000/18001-000/18007-000

Landauer Recall 93524 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-0616-2024
Event ID93524
Event DescriptionLandauer microSTARii reader, Medical Dosimetry System, Model Numbers: 18000-000/18001-000/18007-000
Product TypeDevices
DistributionUS and Australia, Brazil, Canada, China, France, Germany, India, Japan, Mexico, Rep of Korea, Taiwan, Thailand, Viet Nam
Quantity584 units
Recall ReasonMicroSTARii readers may be contributing to the nanodot dosimeter measurement inaccuracy in two possible ways. The first is related to the interaction of the reader s LED beam profile with the nanoDot dosimeter. The second is debris formation due to drawer actuation, resulting in a change of reader response that may not be detectable unless known dose QCs are being run regularly.
Device Classification20240103
Device Code InfoAll serial numbers.
Center Classification Date20231222
Recall Initiation Date20231115
Recalling FirmLandauer