Synthes (USA) Products LLC Recall 89428

Description: Universal Chuck-The Universal Chuck is an Instrument Handle provided as an alternative instrument available for use in the TN-ADVANCED" Tibial Nailing System (TNA), TFN-ADVANCED" Proximal Femoral Nailing System (TFNA), RFN-ADVANCED" Retrograde Femoral Nailing System (RFNA), FRN-ADVANCED" Femoral Recon Nailing System (FRNA), and Flexible Monobloc Reamers. Instrument Handles are intended for holding instruments for manipulation during orthopedic surgery. Product No.: 03.043.001

Synthes (USA) Products LLC Recall 89428 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-0634-2022
Event ID89428
Event DescriptionUniversal Chuck-The Universal Chuck is an Instrument Handle provided as an alternative instrument available for use in the TN-ADVANCED" Tibial Nailing System (TNA), TFN-ADVANCED" Proximal Femoral Nailing System (TFNA), RFN-ADVANCED" Retrograde Femoral Nailing System (RFNA), FRN-ADVANCED" Femoral Recon Nailing System (FRNA), and Flexible Monobloc Reamers. Instrument Handles are intended for holding instruments for manipulation during orthopedic surgery. Product No.: 03.043.001
Product TypeDevices
DistributionUS Nationwide distribution.
Quantity600 units
Recall ReasonTop cap may loosen and detach if the Universal Chuck becomes jammed and the user attempts to manually free the device, internal ball bearings may fall out onto the surgical field and cause surgical delay or, and if not recovered, adverse tissue reaction and infection.
Device Classification20220216
Device Code InfoLot/Serial No.: 20013101, 20052701, 20228901, 20229301, 20253901, 20558602, 20698102, 20777201 21869701, 21983904, 21995301 GTIN: 10886982297372
Center Classification Date20220210
Recall Initiation Date20211222
Recalling FirmSynthes (USA) Products LLC