Percussionaire Corporation Recall 93476

Description: High Frequency Transport Phasitron Breathing Circuit Kit REF: A50605-D, Rx Only, CE 0123

Percussionaire Corporation Recall 93476 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-0637-2024
Event ID93476
Event DescriptionHigh Frequency Transport Phasitron Breathing Circuit Kit REF: A50605-D, Rx Only, CE 0123
Product TypeDevices
DistributionUS:AL AR CA CT FL GA IL IN KS LA MI MN MO NC NJ OH OR TN TX UT VA OUS: None
Quantity2145 units
Recall ReasonDue to an assembly error within the breathing circuit, their pressure output of the breathing circuit kit can be affected and have severe impact to the end user
Device Classification20240124
Device Code InfoUDI: (01)10849436000331/Lot # WO05216, WO05254, WO05321, WO05392, WO05559, WO05588, WO05684, WO05749, WO05896, WO05930,WO06028
Center Classification Date20240117
Recall Initiation Date20231212
Recalling FirmPercussionaire Corporation