Olympus Corporation of the Americas Recall 89404
Description: OER-Elite Endoscope Reprocessor-for use in cleaning and high-level disinfection of heat sensitive Olympus flexible endoscopes, their accessories, and endoscope reprocessor accessories
Olympus Corporation of the Americas Recall 89404 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-0638-2022 |
Event ID | 89404 |
Event Description | OER-Elite Endoscope Reprocessor-for use in cleaning and high-level disinfection of heat sensitive Olympus flexible endoscopes, their accessories, and endoscope reprocessor accessories |
Product Type | Devices |
Distribution | Nationwide |
Quantity | 295 units |
Recall Reason | The detergent and alcohol tank tubing may have been incorrectly assembled in the OER-Elite. If the tubing is incorrectly assembled, alcohol will be dispensed instead of detergent for the cleaning cycle and detergent will be dispensed instead of alcohol for the alcohol cycle, resulting in a ineffective endoscope reprocessing and patient exposure to residual detergent |
Device Classification | 20220223 |
Device Code Info | All serial numbers. UDI: 04953170404047 |
Center Classification Date | 20220211 |
Recall Initiation Date | 20211213 |
Recalling Firm | Olympus Corporation of the Americas |
Initial Notification | Letter |