Product Safety Recalls

Product Recall Tracker

Micro-X Ltd. Recall 95603

Description: MICRO-X Rover Mobile X-ray System, # MXU-RV35

Micro-X Ltd. Recall 95603 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-0664-2025
Event ID95603
Event DescriptionMICRO-X Rover Mobile X-ray System, # MXU-RV35
Product TypeDevices
DistributionWorldwide and US Nationwide Distribution: NJ, PR, CA
Quantity5
Recall ReasonShots were terminated by the mAs integrator; however, the 4 ms exposure time did not allow enough tolerance to achieve diagnostic exposures due to variations in hardware and how different timers are started.
Device Classification20241218
Device Code InfoUDI-DI: 9357123000037. Serial Numbers: 00334, 00348, 00349, 00351, 00361.¿
Center Classification Date20241212
Recall Initiation Date20230531
Recalling FirmMicro-X Ltd.