Becton Dickinson & Company Recall 89486

Description: BD Vacutainer Luer Lok Access Device - intended as a sterile, single use device designed to provide a connection to a catheter port equipped with a threaded female luer during blood or urine sample collection using evacuated BD Vacutainer Blood or Urine Collection Tubes Catalog Number: 364902, 36490200

Becton Dickinson & Company Recall 89486 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-0665-2022
Event ID89486
Event DescriptionBD Vacutainer Luer Lok Access Device - intended as a sterile, single use device designed to provide a connection to a catheter port equipped with a threaded female luer during blood or urine sample collection using evacuated BD Vacutainer Blood or Urine Collection Tubes Catalog Number: 364902, 36490200
Product TypeDevices
DistributionUS Nationwide distribution in the state of IL.
Quantity288,152,790 Units
Recall ReasonMay shed contaminant particles into the urine specimen that may be misidentified as Red Blood Cells when used in conjunction with the Sysmex UF-1000i or UX-2000 Analyzers and lead to false elevations in RBC counts
Device Classification20220302
Device Code InfoAll Lots within Expiry Catalog No: 364902 UDI: (01)50382903649025 Catalog Number: 36490200 UDI:(01)60382903649022
Center Classification Date20220223
Recall Initiation Date20220106
Recalling FirmBecton Dickinson & Company
Initial Notification Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit