Shirakawa Olympus Co., Ltd. Recall 93583

Description: Model No. CYF-V2, VISERA Cysto-Nephro Videoscope

Shirakawa Olympus Co., Ltd. Recall 93583 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-0691-2024
Event ID93583
Event DescriptionModel No. CYF-V2, VISERA Cysto-Nephro Videoscope
Product TypeDevices
DistributionUS Nationwide distribution.
Quantity2210 US; 6002 OUS
Recall ReasonOlympus has updated the instructions for use: Reusable cleaning brushes have been removed from the manual cleaning. Single use brushes remain compatible.
Device Classification20240117
Device Code InfoUDI-DI: 04953170287091 04953170339431 04953170411199 All serial numbers
Center Classification Date20240109
Recall Initiation Date20231204
Recalling FirmShirakawa Olympus Co., Ltd.